BlogSleeping & Neurology

Designing an EEG and PSG Headbox Enclosure

How to lay out, label and mount an EEG or PSG headbox: 10–20 positions printed on the panel, touch-proof socket spacing, strain relief and an RFQ list.

For its size, a headbox carries more connectors than anything else in a sleep or EEG lab. A technologist plugs in twenty or more electrode leads by position, often in dim light, then wipes the box down after the study. This post works through the enclosure in the order we would review it: the input list, the jack layout and its printed map, the sockets, strain relief, mounting and cleaning.

For where the headbox sits among the other equipment in the room, see our overview of enclosures for sleep study equipment.

Start from the input list, not the box

List every input and its connector part number before sketching the housing. A polysomnography headbox typically carries:

  • Scalp EEG. The American Academy of Sleep Medicine (AASM) recommends frontal, central and occipital sites referenced to the opposite mastoid (F4-M1, C4-M1, O2-M1), with F3, C3, O1 and M2 as backups.
  • Eye movement (E1, E2), chin EMG (three electrodes), left and right leg EMG, and ECG.
  • Patient ground and, depending on the amplifier, a common reference.
  • DC or auxiliary inputs for airflow, effort belts, oximetry and body position, usually on other connector types.

A clinical EEG headbox carries all 19 scalp positions of the 10–20 system, references and often extra 10–10 positions. Count the inputs, add the ones you expect in the next product version, and the panel area follows. Adding sockets later is a program edit for us, but a panel sized with spare room avoids moving everything else when you do.

Lay the jacks out on a printed head map

A technologist finds a jack fastest when its place on the panel matches the electrode’s place on the head, which is why headboxes commonly arrange EEG inputs on a head outline. The 10–20 system makes that layout logical. Each position is named by region (Fp frontopolar, F frontal, C central, T temporal, P parietal, O occipital), with odd numbers on the left, even numbers on the right and z on the midline. The name comes from the spacing: adjacent electrodes sit 10 or 20 percent of the front-to-back or side-to-side skull measurement apart.

The map is where a fabricated panel earns its place. We print legends, the head outline and color fields directly on the sheet with UV-cured ink, positioned from the same CAD model that places the cutouts, so a label cannot drift away from its jack. Details worth specifying:

  • The naming convention your software uses. The 10–10 nomenclature renamed T3, T4, T5 and T6 as T7, T8, P7 and P8, and mastoid references (M1, M2) are labeled differently from earlobe references (A1, A2).
  • Color by function. One color for scalp EEG, others for the EOG, EMG and ECG pairs, and a distinct mark for ground.
  • A tactile marker. A raised TEXTUR3D mark beside the ground and reference jacks makes them easy to find by touch in a darkened room.
  • White ink on dark sheet. White prints opaque, so light legends read on black or dark grey ABS without paint.
Light grey housing with curved side panels and a dark front panel set into it, cut for a display window and a row of buttons
A contrasting front panel set into a light housing. A jack map can be printed on a panel like this flat, or after assembly.

UV-printed legends adhere better than silk screen. For a mark that must survive heavy abrasion, engraving is the more permanent choice. The options are compared on our printed enclosures page.

Touch-proof sockets and the panel around them

Electrode leads end in shrouded 1.5 mm touch-proof plugs (the DIN 42802 style), which became standard after leads with exposed pins were plugged into mains outlets. The panel sockets must accept them, and the enclosure decides three things around each one.

  1. Pitch. Space sockets for the grip of the plug, not for the hole. Take the plug’s outside diameter from its data sheet and leave room for a gloved thumb and finger to pull one plug without disturbing its neighbors. Check it on the first prototype before the layout is released.
  2. Hole tolerance. If the sockets are mounted on a PCB behind the panel, the holes locate to the board through the mounting bosses. We hold ±0.2 mm on the flat pattern and ±0.4 mm on features located from the reverse side. Add the board’s own tolerance and size the clearance around each socket body from the total.
  3. Coating clearance. If the housing is EMI coated inside, mask a ring around each input socket so the coating cannot touch a contact. Patient isolation lives in the electronics, and the enclosure must not undo it; enclosure choices that affect signal quality covers grounding the coating.

Put the DC inputs and the oximeter port on a separate area or a separate bonded panel, with connector types that differ from the electrode sockets, so an EEG lead cannot go into a sensor input.

Strain relief for the lead bundle and the trunk cable

Twenty-odd leads hanging from one face add up to a real load, and every time the patient turns over, the pull arrives at the sockets. Tilt the jack face so the bundle falls toward the patient without a sharp bend at the plugs; a bend in the sheet gives an angled face without an extra joint, up to 115 degrees on a standard bend. Then give the bundle a place to be gathered close to the panel, such as machined slots for a hook-and-loop strap.

White benchtop housing with a raised, angled display bezel, a slot below it and a blue side panel
An angled face built from bent sheet. The same geometry tilts a jack panel toward the bed.

The trunk cable to the amplifier needs a real clamp. Bring it in through a bent wall or a screwed clamp on brass inserts, never through a feature that loads a solvent joint in tension, because bonded joints are meant to carry shear and compression. If you already use a strain relief part, send its data sheet and we cut the opening to it.

Mounting at the head of the bed

Headboxes hang on a wall bracket, clamp to a bed rail or pole, lie on the mattress beside the pillow, or go with an ambulatory patient. Decide which before the flat pattern is final, because each needs a different back.

MountingEnclosure featureDesign note
Wall bracketKeyhole slots, or a rear plate on brass insertsLifts off for cleaning without tools
Pole or bed-rail clampRear plate with brass inserts for a standard clampInserts rather than self-tap bosses, since the clamp comes off for service
On the bedRecessed rubber feet, radiused edgesRecessed feet do not slip around over time; 3 mm sheet keeps the weight down
Worn by the patientBelt clip or strap slotsStandard clips cost less than custom features and are easier to service

Self-tap bosses suit screws installed once or twice; anything removed for service needs brass inserts. Keep bosses under 9 mm tall, with at least 4 mm of thread depth for self-tapping screws. Insert sizes are in our functional features guide.

Designing the box to be cleaned

Electrode paste and gel end up on the headbox. Keep the jack face as one flat machined panel and put the joint lines on the sides, where the 0.5 mm chamfer on the tongue edge leaves a seam a wipe can follow. Flush-mounted sockets clean more easily than sockets set into pockets, which collect paste. Legends printed on the part have no label edge to lift and trap gel. Which plastic survives your lab’s disinfectant, and the acetone used to remove collodion, is a materials question covered in choosing materials for patient-side devices.

What to send for a headbox quote

  1. The input list with connector part numbers, including inputs reserved for the next version.
  2. The PCB model if sockets are board-mounted, in STEP, IGES, SAT, SolidWorks, STL or DXF.
  3. The head-map artwork, or the naming convention and colors to print.
  4. The mounting method, trunk cable diameter and strain relief part.
  5. The cleaning agents used in the lab and any flame rating the device needs.
  6. Whether the housing is EMI coated, and where the coating must be masked.
  7. Quantities and the release schedule.

Send it through our CAD upload page and an engineer replies within one business day. You approve a 3D model before any sheet is cut.

Keep reading

Related articles

Start your enclosure

Send the board. We’ll send the box.

Quotes come from an engineer, not a form robot. Attach CAD, photos or a sketch and we reply within one business day.

Request a Quote Call 425-493-1223