Scalp EEG arrives in the tens of microvolts, surface EMG from tens of microvolts to a few millivolts, and ECG at around a millivolt. At those levels the enclosure is part of the measurement chain, but a smaller part than many teams expect. This post sets out which interference a housing can reduce, which it cannot, and how to specify a plastic enclosure so its shielding helps rather than hurts.
Where the noise actually gets in
Most mains interference in a biopotential recording enters through the patient and the leads, not through the walls of the box. The patient’s body couples capacitively to nearby mains wiring and carries a common-mode voltage. A differential amplifier with good common-mode rejection removes most of it until the electrode impedances differ; the mismatch converts part of the common-mode voltage into a differential signal the amplifier cannot tell from EEG. Loops formed by the leads add magnetic pickup on top.
No enclosure fixes a poorly prepared electrode or a lead draped over a power cord. What the housing can do is narrower: shield the high-impedance front end from electric fields and radio, control where openings and cables breach that shield, give the inputs distance from noisy parts, and avoid new problems such as a floating shield or a bridged isolation barrier.
| Source | How it couples | What the enclosure can do | What it cannot do |
|---|---|---|---|
| Mains electric field: wiring, lamps, other equipment | Capacitively, into the front end, the leads and the patient | A grounded conductive coating around the front end | Anything about the leads or the patient |
| Mains magnetic field: transformers, power cords | Induction into loops in the leads and internal wiring | Room to separate the inputs from power parts | Block it; a thin coating has almost no effect at 50 or 60 Hz |
| Radio: phones, Wi-Fi, two-way radios | Rectified in the input stage, showing up as low-frequency artifact | Coating, short apertures, cable shields terminated at entry | Replace RF filtering on the input circuitry |
| Internal switching: DC-DC converters, display backlights | Near-field coupling inside the box | Distance, in a housing sized for it | Correct a poor board layout |
A plastic housing shields only once it is coated
ABS, polycarbonate and acrylic are insulators and contribute nothing to shielding on their own. We apply an EMI and RFI conductive coating to the inside surfaces after the housing is bonded, so it follows the corners and joint lines as a continuous film, and its conductivity is checked before the housing moves on. The outside keeps its sheet color, texture and printing.
Be clear about what that layer does. A thin conductive film works against electric fields and radio frequencies. Against a 50 or 60 Hz magnetic field it does almost nothing, because stopping a low-frequency magnetic field takes thick conductive or high-permeability material. If a transformer or mains wiring sits near the inputs, the answers are distance and small loop areas. Coating options and masking are described on our EMI shielding plastic enclosures page.
Connect the coating to a defined point
An unconnected coating is a floating conductor. It can couple noise from outside to the circuits inside rather than diverting it, and do more harm than bare plastic. Every coated housing needs a deliberate connection.
- How. Heat-staked brass inserts and metal bosses give the coating a fastener point for a ground strap or a board-mounted spring contact. Use brass inserts rather than self-tapping bosses at any screw that carries a ground, so the contact survives the housing being opened for service. Insert options are in our functional features guide.
- Where. The connection point is the electronics engineer’s decision, and in a device with an isolated patient front end it has safety consequences. The coating becomes part of whichever circuit it is tied to, and it must keep the clearances that circuit requires. Tell us the point and we place the insert there.
- Across isolation barriers. The coating follows every inside surface. If one box holds both patient-side circuits and mains- or USB-side circuits, mask or split the coating so it does not run across the barrier or shorten the creepage and clearance distances next to it.
Openings and cable entries
Every opening is a gap in the shield, and how much a gap leaks depends on its longest dimension more than on its area. Many short vent slots leak less than one long slot with the same open area. We machine vents as slots of 2 mm minimum width with the cross supports retained, which keeps each slot short and the wall rigid.
Cables matter more than vents. A cable that enters the box unfiltered carries outside noise straight past the shield. Group connectors on one bonded panel so the largest openings sit in one place, and terminate a shielded trunk cable to the coating or the ground point where it enters, around its full circumference if the connector allows, rather than through a long pigtail. Electrode inputs are different: the leads are part of the measurement, so their RF filtering belongs on the board right behind the sockets. Socket layout for a many-input box is covered in designing an EEG and PSG headbox enclosure.
Keep shielding off antennas, displays and touch areas
A wireless recorder or headbox needs a hole in the shield where the antenna is. Plastic over an antenna is fine; coating over it is not. Mask a keep-out area around the antenna to the module maker’s guidance, and keep inserts and other metal out of it too. The same masking applies to display windows, capacitive touch areas, label recesses and printed faces that must stay non-conductive.
If the display sits near the front end, connect its metal frame, where it has one, to the coating or the ground point. Mark every masked area on the model you send: masking becomes part of the coating operation for every lot, so the unit that passed in the test chamber matches the unit that ships.
Use the size of the housing as a noise control
In a molded product the box is fixed early and the electronics are squeezed into it. In a fabricated housing, adding 20 mm of length to separate the input stage from a switching regulator, a backlight driver or the power entry is a program edit. Distance is the cheapest attenuator for near-field coupling and for magnetic fields from power parts, and a mains-powered bedside unit benefits most. Put the power entry and supply at the end farthest from the electrode inputs, and route internal power wiring away from them.
Test early, then change the housing rather than the board
EMC requirements for medical electrical equipment come from IEC 60601-1-2, and the testing is yours: we do not certify shielding performance. What we control is a repeatable enclosure, with the same program, coating operation and inspection on every lot under our ISO 9001:2015 system. A coated prototype in 3 to 21 working days can go to your pre-compliance bench, and if a slot, a masked area or a ground point has to move, the revision is a program or process edit rather than a tool change. More on how we build housings for clinical devices is on our medical plastic enclosures page.
Send the model with coated faces, masked areas and ground points marked, and an engineer will review it against the coating rules before quoting. Upload your CAD files to start.